User's Manual
Table Of Contents
- Chapter 1
- Beginning Your Dexcom G5® Mobile Continuous Glucose Monitoring (CGM) System Journey
- Chapter 2
- Indications for Use and Safety Statement
- Chapter 3
- Risks and Benefits
- Chapter 4
- What’s the G5 Mobile?
- Chapter 5
- Set Up Your Display Devices
- Chapter 6
- Starting a Sensor Session: Inserting Sensor, AttachingTransmitter, and Starting Your Session
- Chapter 7
- Calibration
- Chapter 8
- Ending Your Sensor Session and Transmitter Session
- Chapter 9
- Home Screen, Rate of Change Arrows, and Errors
- Chapter 10
- Daily Events Affect Your Glucose Trends and Patterns
- Chapter 11
- Sensor Glucose Alarm and Alerts
- Chapter 12
- On the Go With G5: Customizing Your Alarm and Alerts
- EVERYTHING ELSE G5
- Chapter 13
- Warranty: The Fine Print
- Chapter 14
- How to Take Care of Your G5 Mobile
- Chapter 15
- On the Go With G5 Mobile: Getting Through Security
- Chapter 16
- Need Help? You’re Not Alone!
- Chapter 17
- Technical Information
- Chapter 18
- Troubleshooting
- Chapter 19
- Symbols on Package Labels
- SHARING IS CARING
Dexcom G5 Mobile System User Guide
Technical Information
280
Table 5-A. Hypoglycemia Alert and Detection Rate Evaluation in Reference to
YSI 15 Minutes Before and After (Adult)
Hypoglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed
Detection Rate
55
Original
50% 50% 71% 29%
Software 505
71% 29% 68% 32%
60
Original
64% 36% 75% 25%
Software 505
85% 15% 83% 17%
70
Original
79% 21% 83% 17%
Software 505
92% 8% 91% 9%
80
Original
87% 13% 86% 14%
Software 505
95% 5% 90% 10%
90
Original
90% 10% 89% 11%
Software 505
96% 4% 94% 6%
1
Both sets of study data are presented and are labeled as
Original
(SW10050) or
Software 505
(SW10505).