User's Manual

Table Of Contents
Dexcom G5 Mobile System User Guide
Technical Information
280
Table 5-A. Hypoglycemia Alert and Detection Rate Evaluation in Reference to
YSI 15 Minutes Before and After (Adult)
Hypoglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed
Detection Rate
55
Original
50% 50% 71% 29%
Software 505
71% 29% 68% 32%
60
Original
64% 36% 75% 25%
Software 505
85% 15% 83% 17%
70
Original
79% 21% 83% 17%
Software 505
92% 8% 91% 9%
80
Original
87% 13% 86% 14%
Software 505
95% 5% 90% 10%
90
Original
90% 10% 89% 11%
Software 505
96% 4% 94% 6%
1
Both sets of study data are presented and are labeled as
Original
(SW10050) or
Software 505
(SW10505).