User's Manual
Table Of Contents
- Chapter 1
- Beginning Your Dexcom G5® Mobile Continuous Glucose Monitoring (CGM) System Journey
- Chapter 2
- Indications for Use and Safety Statement
- Chapter 3
- Risks and Benefits
- Chapter 4
- What’s the G5 Mobile?
- Chapter 5
- Set Up Your Display Devices
- Chapter 6
- Starting a Sensor Session: Inserting Sensor, AttachingTransmitter, and Starting Your Session
- Chapter 7
- Calibration
- Chapter 8
- Ending Your Sensor Session and Transmitter Session
- Chapter 9
- Home Screen, Rate of Change Arrows, and Errors
- Chapter 10
- Daily Events Affect Your Glucose Trends and Patterns
- Chapter 11
- Sensor Glucose Alarm and Alerts
- Chapter 12
- On the Go With G5: Customizing Your Alarm and Alerts
- EVERYTHING ELSE G5
- Chapter 13
- Warranty: The Fine Print
- Chapter 14
- How to Take Care of Your G5 Mobile
- Chapter 15
- On the Go With G5 Mobile: Getting Through Security
- Chapter 16
- Need Help? You’re Not Alone!
- Chapter 17
- Technical Information
- Chapter 18
- Troubleshooting
- Chapter 19
- Symbols on Package Labels
- SHARING IS CARING
Dexcom G5 Mobile System User Guide
Technical Information
285
Table 6-A. Hyperglycemia Alert and Detection Rate Evaluation in Reference
to YSI 15 Minutes Before and After (Adult)
Hyperglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hyperglycemia
Detection Rate
Hyperglycemia
Missed
Detection Rate
120
Original
95% 5% 98% 2%
Software 505
98% 2% 100% 0%
140
Original
94% 6% 97% 3%
Software 505
97% 3% 99% 1%
180
Original
92% 8% 97% 3%
Software 505
97% 3% 99% 1%
200
Original
92% 8% 97% 3%
Software 505
96% 4% 98% 2%
220
Original
91% 9% 95% 5%
Software 505
94% 6% 98% 2%
240
Original
91% 9% 94% 6%
Software 505
93% 7% 95% 5%
300
Original
82% 18% 86% 14%
Software 505
86% 14% 90% 10%
1
Both sets of study data are presented and are labeled as
Original
(SW10050) or
Software 505
(SW10505).