User's Manual
Table Of Contents
- Chapter 1
- Beginning Your Dexcom G5® Mobile Continuous Glucose Monitoring (CGM) System Journey
- Chapter 2
- Indications for Use and Safety Statement
- Chapter 3
- Risks and Benefits
- Chapter 4
- What’s the G5 Mobile?
- Chapter 5
- Set Up Your Display Devices
- Chapter 6
- Starting a Sensor Session: Inserting Sensor, AttachingTransmitter, and Starting Your Session
- Chapter 7
- Calibration
- Chapter 8
- Ending Your Sensor Session and Transmitter Session
- Chapter 9
- Home Screen, Rate of Change Arrows, and Errors
- Chapter 10
- Daily Events Affect Your Glucose Trends and Patterns
- Chapter 11
- Sensor Glucose Alarm and Alerts
- Chapter 12
- On the Go With G5: Customizing Your Alarm and Alerts
- EVERYTHING ELSE G5
- Chapter 13
- Warranty: The Fine Print
- Chapter 14
- How to Take Care of Your G5 Mobile
- Chapter 15
- On the Go With G5 Mobile: Getting Through Security
- Chapter 16
- Need Help? You’re Not Alone!
- Chapter 17
- Technical Information
- Chapter 18
- Troubleshooting
- Chapter 19
- Symbols on Package Labels
- SHARING IS CARING
Dexcom G5 Mobile System User Guide
Technical Information
291
Table 7-B. Percentage of System Readings
1
Within YSI Values With Data
Stratified in 2-Hour Increments After Calibration (Pediatric)
Time from
Calibration
Study
2
Number of
Paired
CGM-YSI
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
than
40/40%
YSI
0-2 hours
Original
648 65% 75% 87% 7%
Software 505
545
83% 91% 97% 1%
2-4 hours
Original
649 51% 67% 86% 7%
Software 505
460 72% 89% 96% 2%
4-6 hours
Original
630 51% 61% 80% 10%
Software 505
428 77% 88% 95% 2%
6-8 hours
Original
409 52% 68% 85% 5%
Software 505
325 88% 92% 94% 3%
8-10 hours
Original
296 53% 69% 84% 7%
Software 505
305 86% 93% 97% 1%
10-12 hours
Original
253 58% 74% 89% 5%
Software 505
198 89% 94% 98% 0%
12-14 hours
Original
37 32% 38% 65% 22%
Software 505
1 100% 100% 100% 0%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as
Original
(SW10050) or
Software 505
(SW10505).