Technical Information
296
In the
Original Adult
Study, 108 sensors were evaluated. Ninety-four percent
(94%) of the sensors lasted until Day 7 (145-168 hours). There were 6 (6%)
sensors that ended early, four of which lasted more than 3 days.
In the
Software 505 Adult
Study, 51 sensors were evaluated. Ninety-eight
percent (98%) of the sensors lasted until Day 7 (145-168 hours). There was 1
(2%) sensor that ended early, which lasted until day 5 of the sensor wear.
In the
Original Pediatric
Study, 351 sensors were evaluated. Eighty-five
percent (85%) of the sensors lasted until Day 7 (145-168 hours).
In the
Software 505 Pediatric
Study, 77 sensors were evaluated. Ninety-four
percent (94%) of the sensors lasted until Day 7 (145-168 hours).
Table 9-A. Number of Readings Provided by Each Sensor Over 7 Days
(Adult)
% of Total Possible
Readings Provided
Study
1
Total Readings
Provided
(Min-Max)
% of Systems Providing
That Number of Readings
0-25%
Original
167-491 2%
Software 505
0 0%
26-50%
Original
719-914 4%
Software 505
856-856 2%
51-75%
Original
1267-1267 1%
Software 505
1253-1253 2%
76-100%
Original
1811-1992 94%
Software 505
1497-1992 96%
1
Both sets of study data are presented and are labeled as
Original
(SW10050) or
Software 505
(SW10505).