User's Manual

Table Of Contents
Dexcom G5 Mobile System User Guide
Technical Information
308
Transmitter Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test
Level
Transmitter
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Receiver Product Specifications
Part Number
MT22949
Reading Frequency
Every 5 minutes
Dimensions
Length: 4.0 inches
Width: 2.5 inches
Thickness: 0.55 inches
TX/RX Frequencies
2.402-2.480 GHz
Bandwidth
1.22 MHz
Maximum Output Power
2.5 mW EIRP
Modulation
Gaussian Frequency-Shift Keying
Data Rate
1 Mbps