User's Manual
Table Of Contents
- Chapter 1
- Beginning Your Dexcom G5® Mobile Continuous Glucose Monitoring (CGM) System Journey
- Chapter 2
- Indications for Use and Safety Statement
- Chapter 3
- Risks and Benefits
- Chapter 4
- What’s the G5 Mobile?
- Chapter 5
- Set Up Your Display Devices
- Chapter 6
- Starting a Sensor Session: Inserting Sensor, AttachingTransmitter, and Starting Your Session
- Chapter 7
- Calibration
- Chapter 8
- Ending Your Sensor Session and Transmitter Session
- Chapter 9
- Home Screen, Rate of Change Arrows, and Errors
- Chapter 10
- Daily Events Affect Your Glucose Trends and Patterns
- Chapter 11
- Sensor Glucose Alarm and Alerts
- Chapter 12
- On the Go With G5: Customizing Your Alarm and Alerts
- EVERYTHING ELSE G5
- Chapter 13
- Warranty: The Fine Print
- Chapter 14
- How to Take Care of Your G5 Mobile
- Chapter 15
- On the Go With G5 Mobile: Getting Through Security
- Chapter 16
- Need Help? You’re Not Alone!
- Chapter 17
- Technical Information
- Chapter 18
- Troubleshooting
- Chapter 19
- Symbols on Package Labels
- SHARING IS CARING
Dexcom G5 Mobile System User Guide
Calibration
99
6.9 Receiver
Bluetooth
Tips
Your transmitter and receiver begin communicating once you start a sensor
session. After approximately 30 minutes, if the
Bluetooth
symbol is solid, and
not blinking, your transmitter and receiver are talking to each other.
• If blinking,
Bluetooth
is looking for your transmitter
o Make sure your transmitter and receiver are within 20 feet of
each other
o Don’t start a sensor session until they are paired.
If the
Bluetooth
icon is blinking and
Signal Loss
displays, your transmitter and
receiver are not communicating.
No Communication Between Transmitter and Receiver
Step
What you see
What you do
1
Tap OK.
Verify display device and transmitter are
within 20 feet of each other without
obstruction.
Wait up to 30 minutes.
Don’t calibrate. Use BG meter for BG
reading.
No sensor glucose readings display until
error is fixed.
More than 30 minutes? Contact Technical
Support (available 24/7) at:
• TechSupport@dexcom.com
• Toll free:
1.888.738.3646
• Toll call:
1.858.200.0200