Instruction Manual Type K Thermocouple Thermometer 68X243624 Rev. 0 02/04 Technology Made Easy ...
Preface This manual serves to explain the use of the Digi-Sense® Type K thermocouple thermometer. This manual functions in two ways: first, as a step by step guide to help you operate the meter; second, it serves as a handy reference guide. This manual is written to cover as many anticipated applications of the Type K meter as possible. If there are doubts in the use of the Type K meter, do not hesitate to contact the nearest Eutech Instruments/ Oakton Instruments Authorized Distributor.
Table of Contents 1. INTRODUCTION 1 2. GETTING STARTED 2.1 Description of Keypad Functions 2.2 Description of LCD Annunciators 2.3 Inserting & Removing the Rubber Boot / Stand (Optional Purchase) 2.4 Inserting New Batteries 2.5 Battery Replacement 2.6 Connecting the Temperature Sensor 2.7 Switching the Meter On 2 2 2 3 3 4 4 4 3. CALIBRATION 3.1 Temperature Calibration 5 5 4. MEASUREMENT 4.1 Taking Measurements 4.2 Holding a Reading 4.3 Releasing a Held Reading 7 7 7 7 5.
Instruction Manual Type K Thermocouple Thermometer 1. INTRODUCTION Thank you for purchasing the Digi-Sense Type K thermocouple thermometer. This economical microprocessor-based handheld meter works with any type K probe with color-coded yellow ANSI miniconnectors.
Instruction Manual Type K Thermocouple Thermometer 2. GETTING STARTED 2.1 Description of Keypad Functions The meter has three keys on its splash-proof keypad. These keys include ON / o o OFF, HOLD, and C/ F keys. ON/OFF Powers meter on and shuts the unit off. Meter directly enters measurement mode when you turn it on. o Switches between C and F measurement mode o C/ F o o Activates the calibration mode when used with the ON/OFF key. Decrease the calibration value during the calibration mode.
Instruction Manual Type K Thermocouple Thermometer 2.3 Inserting & Removing the Rubber Boot / Stand (Optional Purchase) 1. To remove meter from rubber boot, push out from the bottom edges of meter until it is completely out of boot. Ensure that the connector of temperature sensor is not connected. 2. To insert meter into rubber boot, slide in from the top of meter before pushing the bottom edges of meter down to set it into position.
Instruction Manual Type K Thermocouple Thermometer 2.5 Battery Replacement A “LO” annunciator in the LCD alerts you when battery power is running low. Replace with the same type as recommended by the manufacturer. K LO o C Figure 5: Low battery indicator Caution: Power off the meter when changing battery. 2.6 Connecting the Temperature Sensor Insert the yellow 2-pin mini connector plug of the temperature sensor into the connector socket on the side of the meter.
Instruction Manual Type K Thermocouple Thermometer 3. CALIBRATION 3.1 Temperature Calibration Your meter is factory calibrated. The meter allows you to enter a 1-point calibration as fine adjustment by changing its offset value. This is useful when using a new probe for the first time. 1. Connect the temperature probe and switch on the meter. 2. Dip the probe in a constant temperature bath, or in liquid whose temperature can be checked against another accurate thermometer.
Instruction Manual Type K Thermocouple Thermometer 7. To confirm value, wait for 5 seconds without any key press. The “CO” display appears signifying conformation and meter goes back to the measurement mode. 8. To abort calibration, press ON/OFF key to power off the meter. NOTE: The offset adjustment window is ±10°C (±18°F) of the factory default value.
Instruction Manual Type K Thermocouple Thermometer 4. MEASUREMENT 4.1 Taking Measurements K °C 1. The meter automatically enters Temperature mode when powered on. The “K” annunciator will be displayed. o o 2. The C or F annunciator displayed in your LCD indicates which mode you are taking measurements in. 3. Press the °C/°F key to toggle between temperature units. 4.2 Holding a Reading To freeze or hold your reading, press HOLD key once.
Instruction Manual Type K Thermocouple Thermometer 5. PROBE CARE AND MAINTENANCE For best results, maintain the thermocouple probe as per the manufacturers instructions. To remove the probe, simply hold firmly onto the plastic yellow mini-connector and pull out of the meter’s socket. Store both the probe and meter into its original packaging when not in use. 6. TROUBLESHOOTING Problem Cause Solution No display when turned on. a) Batteries not in place. a) Re-insert batteries in correct polarity.
Instruction Manual Type K Thermocouple Thermometer 7. SPECIFICATIONS Temperature Range -250 to 1372 °C (-418 to 2501 °F) Type K Resolution t<-99.0 °C 1 °C (1 °F) -99.9 °C299.9 °C 1 °C (1 °F) Accuracy t<-99.9 °C ± 0.25% of reading + 1 °C (2 °F) t>-99.9 °C ±0.25% of reading + 0.5 °C (0.
Instruction Manual Type K Thermocouple Thermometer 8. ACCESSORIES Item Description Eutech Instruments Oakton Instruments Ordering Code Ordering Code 93000-00 93000-00 Type K general purpose probe (immersion into liquids), 5-ft coiled cable. EC-TPGLPK-01M 08516-55 Type K penetration probe (meat, semi-soft materials), 5-ft coiled cable. EC-TPPENK-01M 08516-65 Type K surface probe (direct contact on hot surface, no liquids), 5-ft coiled cable.
Instruction Manual Type K Thermocouple Thermometer 9. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge.
Instruction Manual Type K Thermocouple Thermometer 10. RETURN OF ITEMS Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Goods Authorization” (RGA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss.
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below: Oakton Instruments Eutech Instruments Pte Ltd. P.O Box 5136, Blk 55, Ayer Rajah Crescent, Vernon Hills, IL 60061, USA #04-16/24 Singapore 139949 Tel: (1) 888-462-5866 Tel: (65) 6778 6876 Fax: (1) 847-247-2984 Fax: (65) 6773 0836 E-mail: info@4oakton.com E-mail: marketing@eutechinst.com Web-sites: Web-site: www.eutechinst.com www.4oakton.com www.