Specifications

Furnace
The furnace chamber is heated by four electric resistance
heaters which are embedded in a refractory material. The
chamber is insulated with a ceramic fiber insulation. The
control is located under the furnace chamber and is well
insulated from the heat generated in the furnace chamber.
A door safety switch removes power to the heating ele-
ments whenever the furnace door is opened. The choices
of temperature controllers are as follows:
Temperature Controller: “Basic
Automatic”
The basic automatic control provides simple automatic
control to one set point temperature. A single digital dis-
play will indicate either furnace chamber temperature or
set point temperature. This controller does not provide
over temperature protection.
Temperature Controller:
“Automatic”
This furnace controller consists of a microprocessor
based three mode temperature controller having adjust-
able over temperature protection and appropriate output
switching devices to control the furnace.
Temperature Controller:
“Programmable”
This furnace controller consists of a microprocessor
based three mode temperature controller/programmer
with adjustable over temperature protection and appropri-
ate output switching devices to control the furnace.
6
Principles of Operation
Note
When in the program RUN mode, the
programmer/controller serves to pro-
vide a programmed temperature pro-
file as described earlier. When in the
AUTOMATIC mode, the unit serves
as a single set point automatic tem-
perature controller.