Operating instructions
G3 Series Malfunctions
Error messages
6-15
000012,03,02-2009, jmu
6
0x00017103 Temperature too high in the electronics unit.
The system may be damaged.
Temperature %0s °C.
The temperatu re in the electronics unit exceeded a c ritical threshold . Check that the fan is n ot covered. Ch eck the fan fil ter and replac e it if it is co ntaminated.
382
Cause
The temperature in the BF board exceeded a critical threshold. This may cause
damage to the hardware.
Solution
Switch machine off and leave to cool down. Contact service if this error occurs again.
0x00017301 Command currently not possible.
Instruction was not ready. Try again.
385
Cause
An instruction waits for commands, but is not able to process a prematurely received
command.
Solution
Try another time. Contact the service department
0x00017402 Undefined status of an instruction.
Programming error.
386
Contact service
0x00017502 Error in the slide vacuum.
No help available
0x00017704 Communication error on an internal bus (CAN).
No help available
389
Contact service
0x00017803 Incorrect slide vacuum parameter. Error index %0s
Vacuum slider parameter erro r.
0x00017A03 Vacuum slider current limitation.
No help available
0x00017B02 CAN message has been lost.
An internal communication problem occurred.
393
Contact service
0x00017C03 Incorrect module initialisation Status
Internal softwa re error.
394
Contact service
0x00017D04 Operating unit has lost the communication to the MC.
CAN Tx error.
No help available
395
Contact service
0x00017E04 Operating unit has lost the communication to the MC.
CAN Rx error.
No help available
396
Contact service
0x00017F03 Error when looking for the zero mark of the T axis.
No help available
397
Contact service