User Manual

Floor / Pallet Scale - 5 -
2.3 CONNECTING LOAD CELL TO INDICATOR
1. The 9-pin socket is used to hook up load cell, please refer to graph 2-3.
2. The 4-pin shielded cable is used, and the indicator does not have the function of long distance
compensation.
3. Indicator must be reliably connected to Load cell and shielded-cable of load cell must be reliably
grounded.
4. Sensor and indicator are static sensitive devices. Soldering operation is prohibited.
Port of Indicator Port of load cell
- V Shielded +V Excitation Voltage+
- V Excitation Voltage-
+IN Output Signal+
-IN Output Signal -
+V -IN +IN shielded
(Graph 2-3) Drawing of connecting of load cell
Chapter 3 Operation
3.1 POWER ON AND AUTO ZERO-SETTING
3.1.1 The indicator will perform countdown self-checking when turned on. Then it will enter weighing
mode.
3.1.2 When powered on, if loading weight on the scale deviates from the zero point, but still within
zero set range, the indicator will set zero automatically; if it is out of range, it is necessary to
adjust the zero point or recalibrate or reset.
3.1.3 Press UNIT button to select your weighing unit. There is indication on the readout panel
showing the current weighing mode.
3.2 MANUAL ZERO SETTING
3.2.1 In weighing mode, you can press [Zero] anytime to zero the readout.
3.2.2 If the displayed value deviates from zero point, but still within zero-range, pressing [Zero] key is
available. Otherwise, [Zero] key is invalid. (In this status, please recalibrate or reset zero
parameters)
3.2.3 Zero function works only when the STAB sign is on.
3.3 TARE FUNCTION
When Indicator is in weighing mode, press the [Tare] key, the indicator will deduct the displayed weight
value as tare weight. Then indicator displays net weight as “0”, and Tare sign is on.