Appendix A
Error information:
display
mode
description
□□□□□□
weighing mode
overload or without ADC input
ERR0
H17 Span Calib.
The calibration test weight or the
adjusted keyed-in weight is larger
than the full capacity of the scale.
Change the calibration test weight
or check the input data.
ERR1
H17 Span Calib.
The calibration test weight or the
adjusted keyed-in weight is smaller
than 1% of the full capacity of the
scale. Change the calibration test
weight or check the input data.
ERR2
H17 Span Calib.
The internal resolution of the scale
is not high enough to accept the
calibration value.
ERR3
all modes
One or more parameters had lost.
ERR4
all modes
Rom writing wrong. The indicator
needs to repair.
ERR5
H20 Key-in Span Calib.
If a value of zero is entered, the
indicator will briefly show "Err 5".
ERR9
weighing mode
The span calib. parameters had
been lost. The indicator needs to
reset H17.