Service manual

Rev. 01 Model 515B/C
Service Manual
33
8
Functional Test
The functional tests verify the overall functional integrity of the monitor and sensor. If the
monitor and sensor do not pass these tests, remove from use and contact the Novametrix Service
Department for repair/replacement assistance.
8.1
Equipment required
Line cord (supplied with unit) PN: 600026
SuperBright™ Sensor (Finger sensor-PN: 8744, or Y-Sensor-PN: 8793) with OxySnap™
extension cable-PN: 8853 (8 ft), 8898 (12 ft), or SuperBright™ Sensor (Finger sensor-
8776, or Y-Sensor-PN: 8791).
If using the Y-Sensor it is recommended that one of the Y-Strip tapes be used also. This
will facilitate easy application of the sensor for testing purposes.
Finger Sensor Cat No. 8744
Y-Sensor Cat No. 8793
OxySnap extension cable
Finger Sensor Cat No. 8776
Cat No. 8853