User`s guide
AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 10
3. Equipotential Stud. Earth Ground connector.
4. Power Cord Connection. Port for connecting the RF Generator to power outlet
via the power cord.
5. Fuse. To replace fuse see maintenance section.
6. Power Switch. Toggling this switch turns the RF Generator ON (starting the self-
test) and OFF.
CAUTION: Only modules supplied by AngioDynamics, Inc. should be plugged into the
generator. Plugging in other modules may cause severe damage to the generator.
5. INSTRUCTIONS FOR USE OF THE GENERATOR
5.1 Steps in the Procedure
5.1.1 Patient Preparation
• Apply the Dispersive Electrode pad(s) according to the accompanying instructions for
use and the figures in the Body Atlas (see Dispersive Electrodes Instructions for
Use). The RITA
Dispersive Electrodes must not be used with bipolar devices such as
the Habib
4X. Consult each device’s IFU for more information. The RITA
Dispersive
Electrodes also must only be used with the RF Generator. The entire surface area of
the Dispersive Electrode must be in contact with the patient. Be sure to follow the
package instructions carefully.
• Prepare the patient using the standard technique for electrosurgery. The patient’s entire
body, including extremities, should not come into contact with metal parts which are
earth grounded or which have an appreciable capacitance to earth (for example
operating table supports, etc.). Sufficient layers of electrically insulated sheets should
be used for this purpose. A waterproof cover should be placed over the insulating
sheets, with absorbent sheets placed between the patient and the waterproof cover to
absorb any moisture.
• Skin-to-skin contact, such as between the torso and the arms, and between the legs of
the patient should be avoided by insulating these contacts with sheets or dry gauze.
• Any monitoring electrodes should be placed as far as possible from the Device, and
should incorporate high-frequency current limiting devices. Cables connected to the
Device should not contact the patient or other electrical leads. Needle monitoring
electrodes are not recommended.
• Low power output or failure of the RF Generator to deliver RF energy may indicate faulty
application or connection of the Dispersive Electrode.