User`s guide

AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 11
5.1.2 Setting up the RF Generator
If the Device requires a Main Cable (see Device’s Instructions for Use), sterilize the Main
Cable in accordance with the Instructions for Use accompanying the cable. Ensure that
the cable interconnections are clean and dry prior to use.
Connect the Foot Pedal to the RF Generator.
Connect the Dispersive Electrodes to the RF Generator at the RETURN port.
Turn the RF Generator on, using the switch on the rear panel. (If the RF Generator fails
the self-test, turn the RF Generator off and then turn it on again. If it fails again, call your
local sales representative/distributor or Customer Service at AngioDynamics, Inc.)
When the Generator finishes it self-test, press the RF button to take the Generator out of
Standby Mode.
5.1.3 Programming the RF Generator and Connecting the Devices/Accessories
Set the Control Mode by pressing the “CONTROL MODE” button. (Note: Some
devices only have one mode. See Device’s Instructions for Use).
Set Maximum Power using the “SET POWER (W)” arrow buttons. Set the power as
low as possible for intended purpose.
Inspect the Device.
If the Device does not already have its own main cable, connect the Device to the
Main Cable and pass the other end of the cable from the sterile field to the RF
Generator. The end of the cable that has a flag with a picture of the RF Generator is the
connector that goes to the RF Generator. The other end connects to the Device.
Connect the Main Cable to the RF Generator.
Ensure that all thermocouples are reading approximately the same room
temperature.
If using the Automated Infusion Mode, connect the pump to port B (RS-232 connection)
on the RF Generator and the RS-232 connection on the pump. Ensure the pump is on
and that purging (as described above) is complete.
Disconnect the Device from the Main Cable for placement in the target area
(optional).
Retract the array of electrodes (hooks) of the Device.
Place the Device in the target area according to the instructions with the Device.
Deploy the Device array electrodes according to the instructions with the Device
and/or the RF Generator LCD display.
Reconnect the Device to the Main Cable (if disconnected for placement).
5.1.4 Operation of the RF Generator During a Procedure
Check the LCD display. It should display, “TURN ON (RF) TO START”
Check all displays to confirm the settings and to confirm that the temperature
sensors are functioning properly.