AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 18
3) Remove the fuse from the fuse module
and replace it with the type of fuse listed
under fuses in the specifications section
below.
4) Insert the fuse module back and close
the Line Input module access door.
8.6 Calibration Verification
There are no user calibration adjustments on the Model 1500X RF Generator.
The RITA
1500X RF Generator has been tested and is compliant with
the requirements of the following standards:
- UL60601-1, Standard for Safety Medical Electrical Equipment: General Requirements for
Safety
- Canadian CAN/CSA C22.2 No.601-1, Medical Electrical Equipment: General Requirements for
Safety
- IEC 60601-1, Medical Electrical Equipment: General requirements for basic safety and
essential performance
- IEC 60601-1-2, Medical Electrical Equipment: Electromagnetic compatibility – Requirements
and test
- IEC 60601-2-2, Medical Electrical Equipment: Particular requirements for the safety of high
frequency surgical equipment
In addition to compliance to the above listed standards AngioDynamics, Inc. tests each
generator per a comprehensive test protocol prior to shipment.
The 1500X RF Generator requires no preventative maintenance, calibration, or testing
prior to use. Once the generator is turned on the software runs through a self-test
verifying the functionality of the generator. If the generator fails the self-test, call
Customer Service at AngioDynamics, Inc. – U.S. customers (800) 772-6446;
International customers/distributors +32-2 252 12 02.