User`s guide
AngioDynamics, Inc. Model 1500X User’s Guide and Service Manual
160-103990 Rev 03 Page 27
During use of the RF Generator, the following tests are conducted. If a test fails, the
following message will appear on the LCD:
LCD Display
S
olution
“SYSTEM FAILURE 7: WATCHDOG
TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service.
“SYSTEM FAILURE 3: POWER SUPPLY
TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service.
“SYSTEM FAILURE 8: LOOP OVERRUN
TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service.
“SYSTEM FAILURE 9: TIME ERROR”
“TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service.
“SYSTEM FAILURE 6: TEMP REF
TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service.
“SYSTEM FAILURE 10: SYSTEM ON TOO
LONG
TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service. Note: The
Generator should not be left on, because it is
important that before every ablation that it runs its
self test, which only occurs at startup.
“SYSTEM ERROR 11: SPURIOUS INTERRUPT
TURN POWER OFF”
Turn Power off. Turn Power back on. If error
persists, contact Customer Service.
10.1.3 Impede-Out Recovery Routine
When an impede-out occurs, the Generator attempts to reestablish tissue connection by turning the
RF energy off and infusing saline at a higher rate. While this is occurring, the LCD display will read
“INFUSING – POWER ON in XX” where “XX” is a countdown for when the RF energy will be turned
back on. If there are successive impede-outs, more time and higher infusion is used to establish a
better tissue connection.
10.1.4 Device Troubleshooting
For details regarding a specific Device’s troubleshooting, please see that Device’s Instructions for
Use.