Operating instructions

51
9Maintenanceandfaultrectication
VEGAMET 625 • Double channel HART
28970-EN-150129
Error code Cause Rectication
E053 Sensor measuring
range is not read cor-
rectly
Communication error: Check sensor
cable and screening
E062 Pulse priority too
small
Increaseunder"Output"theentry
"Pulse output all"sothatmax.one
pulse per second is outputted.
E110 Relay switching
points too close to-
gether
Increasethedierencebetweenthe
two relay switching points
E111 Relay switching
points interchanged
Change relay switching points for
"On/O"
E115 Several relays are as-
signef to the pump
control which are not
set to the same fail-
ure mode
All relays which are assigned to the
pump control must be set to the
same failure mode
E116 Several relays that
arenotcongured
with the same mode
are assigned to the
pump control
All relays which are assigned to the
pump control must be set to the
same mode
Depending on the reason for the fault and the measures taken, the
stepsdescribedinchapter"Set up"mayhavetobecarriedoutagain.
9.3 Howtoproceedifarepairisneeded
Youcanndarepairformaswellasdetailedinformationonhowto
proceed at www.vega.com/downloadsand"Forms and certicates".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Printandlloutoneformperinstrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
Please contact the agency serving you to get the address for
thereturnshipment.Youcanndtheagencyonourhomepage
www.vega.com.
Reactionafterfaultrecti-
cation