Specifications

PN 718113AF
1
Laser Induced Fluorescence Detection
Specifications
Table 1 LIF Detection Specifications
Specification Value
Relative Fluorescence Units (RFU) range 0 to 1000 RFU
Dynamic Range (at a dynamic range setting of 1000)
> 10
4
Sensitivity
1 ´ 10
-11
M Sodium Fluorescein with a
signal-to-noise ratio ≥ 2
Baseline Noise
< 0.005 RFU peak to peak
a
Baseline Drift
< 0.2 RFU per hour
a
488 nm Laser Module Dimensions
Height 25.00 in. (63.5 cm)
Width 10.25 in. (26.3 cm)
Depth 14.25 in. (36.0 cm)
635 nm Laser Module Dimensions (Discontinued)
Height 7.5 in. (19.2 cm)
Width 5.25 in. (13.2 cm)
Depth 7.0 in. (17.7 cm)
Wavelength Ranges (for optics)
Excitation 300 to 700 nm
Emission 350 to 750 nm
Filters (optional)
520 nm bandpass filter
488 nm notch filter
675 nm long pass filter
Fiber cable length 6 feet (1.83 m)
Software compatibility Instrument software, Version 2.0 or later
Laser (optional) 3 mW 488 nm (air-cooled) Argon Ion Laser
635 nm Diode Laser