Specifications

PN 718113AF
14
Laser Induced Fluorescence Detection
Installation
Installing the LIF Detector into the Instrument
WARNING
Be sure to power down the instrument before proceeding with component
installation.
Hold the LIF Detector level and place it into the instrument, as shown in Figure 7, keeping the two
female connectors on the LIF Detector level to the two male connectors located on the instrument.
Secure the Detector Module in place by fastening the captive screws to finger tightness.
Preparing the Cartridge for LIF Usage
Build the cartridge as described in the Installation and Maintenance Manual, up to the point when
the aperture plug is installed. The LIF Detector uses a special plug, which is not interchangeable
with those used on the UV/PDA system.
To install the LIF plug and probe stabilizer (see Figure 8):
1
Carefully install the metal LIF plug from the back of the cartridge.
2
Snap on the probe stabilizer such that the lock pins snap into the dimples on the plug.