10 1200 Series Variable Wavelength Detector User Manual
1 Introduction to the Variable Wavelength Detector
Introduction to the Detector
The Agilent 1200 Series variable wavelength detector is designed for highest
optical performance, GLP compliance and easy maintenance with:
higher data rate (27/55Hz) for fast-HPLC with G1314C VWD-SL, see “Set the
Detector Parameters" on page 69,
deuterium lamp for highest intensity and lowest detection limit over a
wavelength range of 190 to 600 nm,
optional flow-cell cartridges (standard 10 mm 14 µl, high pressure 10 mm
1l, micro 5mml, semi-micro 6mml) are available and can be used
depending on the application needs,
easy front access to lamp and flow cell for fast replacement, and
built-in holmium oxide filter for fast wavelength accuracy verification.
For specifications “Performance Specifications" on page 25.
Two version of the Agilent 1200 Series variable wavelength detector are
available:
G1314B VWD 1200 Series Variable Wavelength Detector
standard version
G1314C VWD-SL 1200 Series Variable Wavelength Detector SL
high data rates for fast HPLC
NOTE
The G1314C VWD-SL can be operated with a G1323B Control Module just in standard mode
as G1314B - no higher data rate selection is available.