2 Site Requirements and Specifications
Physical Specifications
Table 2 Physical Specifications
Type Specification Comments
Weight 11 kg
25 lbs
Dimensions
(height × width × depth)
140 × 345 × 435 mm
5.5×13.5×17inches
Line voltage 100 – 240 VAC, ± 10% Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5%
Power consumption 220 VA, 85 W / 290 BTU Maximum
Ambient operating temperature 0–55 °C (32–131 °F)
Ambient non-operating temperature -40–70 °C (-4–158 °F)
Humidity < 95%, at 25–40 °C (77–104 °F) Non-condensing
Operating altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the instrument
Safety standards: IEC, CSA, UL, EN Installation Category II, Pollution Degree 2.