1200 Series Variable Wavelength Detector User Manual 55
Using the Detector 4
Special Settings of the Detector
In this chapter special settings of the G1314B VWD and G1314C VWD-SL are
described (based on the Agilent ChemStation B.02.01).
Control Settings
Figure 20 Detector control settings
•Lamp: turn on and off of
UV-lamp.
•At Power On: automatic lamp-on
at power on.
• Error Method: take error method
or current method (in case of an
error).
• Analog Output Range: can be
set to either 100 mV or 1 V full
scale, see also “Analog Output
Settings" on page 59.
• Lamp Type: can be set to either
G1314-60100 (standard VWD
lamp) or 2140-0590 (DAD lamps),
see also “Exchanging a
Lamp" on page 85.
• Automatic Turn On: lamps can
be programmed (detector must
be on for this).
•Help: online help.