92 1200 Series Variable Wavelength Detector User Manual
8Maintenance
Next steps Reconnect the capillaries, see “Exchanging a Flow Cell" on page 87.
Perform a leak test. If OK, insert the flow cell.
Perform “Wavelength Verification/Calibration" on page 104 to check the
correct positioning of the flow cell.
Replace the front cover.