1200 Series Variable Wavelength Detector User Manual 93
Maintenance 8
Using the Cuvette Holder
This cuvette holder can be placed instead of a flow cell in the variable
wavelength detector. Standard cuvettes with standards in it, for example,
National Institute of Standards & Technology (NIST) holmium oxide solution
standard, can be fixed in it.
This can be used for wavelength verifications.
When required: If your own standard should be used to checkout the instrument.
Tools required: None
Parts required: Cuvette Holder G1314-60200
Cuvette with the “standard”, e.g. NIST certified holmium oxide sample
Figure 31 Cuvette Holder