94 1200 Series Variable Wavelength Detector User Manual
8Maintenance
Preparation for this procedure are:
Remove the normal flow cell.
Have cuvette with standard available.
1 Locate the cuvette holder on the desk.
2 Unscrew the bracket. 3 Insert the cuvette with the sample into the holder. The
clear side of the cuvette must be visible.
Next steps:
Reset the lamp counter as described in the User Interface documentation.
Turn the lamp ON.
Give the lamp more than 10 minutes to warm-up.
Perform “Wavelength Verification/Calibration" on page 104to check the correct positioning of the lamp.
Light path
Clear side