96 1200 Series Variable Wavelength Detector User Manual
8Maintenance
Correcting Leaks
1 Remove the front cover.
2 Use tissue to dry the leak sensor area.
3 Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4 Replace the front cover.
When required If a leakage has occurred in the flow cell area or at the capillary connections.
Tools required Tissue
Two 1/4 inch wrenches for capillary connections
Parts required None
Figure 32 Drying the Leak Sensor
Waste outlet
Leak plane
Leak sensor assembly