User manual

1200 Series Variable Wavelength Detector User Manual 97
Maintenance 8
Replacing Leak Handling System Parts
1 Remove the front cover to have access to the leak handling system.
2 Pull the leak funnel out of the leak funnel holder.
3 Pull the leak funnel with the tubing out of its location.
4 Replace the leak funnel and/or the tubing.
5 Insert the leak funnel with the tubing in its position.
6 Insert the leak funnel into the leak funnel holder.
7 Replace the front cover.
When required If the parts are corroded or broken.
Tools required None
Parts required Leak funnel 5061-3356
Leak funnel holder 5041-8389
Leak tubing (120 mm) 0890-1711
Figure 33 Replacing Waste Handling System Parts
Waste outlet
Leak plan
Leak sensor assembly
Leak tubing
Leak funnel
Leak funnel holder