7.3.2ADJUSTMENTOFCHEMICALSDEFICIENCYSENSORS
POTENTIOMETER:
Graduations:1to9
Sensitivity:1(nonsensitivechemicalsensing/
sensitivefailureindication
9(sensitivechemicalsensing/
nonsensitivefailureindication)
Basicsetting:3
Duetothephysicalpropertiesofrinseaid(e.g.wetting),evensmallestrinseaidquantitiesinsidethehosewillbe
detectedbythedeficiencysensor.Ifthesensorisadjustedtoosensitively,maybedeficiencywillnotbereleased.
TEST"DEFICIENCY"
Flushthesuctionhosethoroughlywithfreshwater
toremoveanychemicals.
Whenthehoseisdrained,therespectiveLEDshouldbe"OF".
TEST"FULL"
Fillthesuctionhose(seechapter5,page8).
TherespectiveLEDshouldlightup.Ifnot,adjustpotentiometeruntiltheLEDlightsup.
Nowthehoseshouldbecompletelyfilledandwithout
airbubbles.
TESTINGTHEPCB
SelectServiceMode(seechapter111.1.2,page28).
Hosesareemptyanddeficiencysensorpotentiometersturnedtoleftstop:
Switchingfunctions"S07"(detergent)and"S08"(rinseaid)mustbe"0".NosensorLEDlightsup.
Potentiometersturnedtoright
stop:
Switchingfunctions"S07"(detergent)and"S08"(rinseaid)mustbe"1".TheLEDoftherespectivecircuitlights
up.

Detergentdeficiencyindication"‐‐0"Rinseaidhosefilled"‐‐1"
Aftertesting:
Setpotentiometers(detergentandrinseaid)tovalue"3"(basedon
testswiththemostcommonproducts).
Detergent
Rinse aid