Operating instructions

When data are output without decimals, the decimal point is suppressed
(except when a certain display mode is selected).
12345678910111213141516
Data output example: +1501.117 mg
12345678910111213141516
Characters:
1st Plus or minus sign or space
2nd–10th Weight with a decimal point; leading zeros = space
11th Space
12th–14th Unit symbol or space
15th Carriage return
16th Line feed
If the weighing system has not stabilized, no unit symbol is output.
For the display format parameter “last numeral off or only at stability” (blank last
numeral by a rounding factor of 10 or display last numeral when the readout
is stable), the 10th character in the string is a space. The reason: the last numeral
without a decimal point is not blanked. Instead, it is given a definite zero,
otherwise the display would go completely blank when the balance is unloaded.
In this case, you will obtain the standard display format.
* = space, E = unit
+
*
1501 . 117
*
mg
*
CR LF
+
******
––––––––––––––––––––––––
***
**
10
7
10
6
10
5
10
4
10
3
10
2
10
1
10
0
*
CR LF
–––––––––––––––––––––––– E E E
0000000
4–4