Operating instructions

Model MC 5 SC 2
Weighing capacity g 5.1 2.1
Readability µg 1 0.1
Tare range (subtractive) g –5.1 –2.1
Repeatability (standard deviation)* µg 1 0.25
Linearity µg ±4 ±0.9
Linearity within 500 mg µg ±2 ±0.4
Response time (average) s –––––––––––––––– 10 ––––––––––––––––
Response time in the IQ-mode (0.01%) s –––––––––––––––– 2.5 ––––––––––––––––
Adaptation to ambient conditions
and application requirements – By selection of 1 of 4 optimized filter levels
Display update (depends on the filter
level selected) s 0.1–0.4 0.2–0.4
Operating temperature range K ––––––– 278...313 (+5 ... +40°C) –––––––
Sensitivity drift within 278 ... 313K /K ––––––––––––––– 1·10
–6
––––––––––––––
Deviation of the readout
when balance is tilted 1 : 1,000 u
ı
g 5 3
Pan size mm Ø 30 Ø 20
Draft shield height mm –––––––––––––––– 122 ––––––––––––––––
Net weight, approx.
– Weighing cell kg –––––––––––––––––3.3–––––––––––––––––
– Evaluation system kg –––––––––––––––––3.1 –––––––––––––––––
Power requirements V~ TNG6 power supply,
230 or 115, +15%...–20%
Frequency Hz ––––––––––––––– 50–60 –––––––––––––––
Power consumption incl. power supply, approx. VA ––––––18.7: maximum; 11: average ––––––
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, gr, dwt,
mg, /lb, tlc, mom, K, tol, bat, MS
Application programs Mass unit conversion; “EUREKA” air buoyancy
correction; differential weighing and
backweighing; diameter determination; tare
memory; net total; weighing in percent;
over/under checkweighing (classification and
filling); counting
Automatic zero tracking device,
can be turned off by menu code –––––––––––––– Standard ––––––––––––––
Built-in interface RS-232 C-S/V24-V28, RS-423/V 10; 7-bit;
parity: even, mark, odd, space; transmission
rates: 150...19,200 baud; 1or 2 stop bits;
software/hardware handshake
* = Standard deviation of the reproducibility according to DIN 1319, Part 3
5–3
▲▲