Model MC 5-0CE SC 2-0CE
Type ––––––––––––––– KC BA 100 –––––––––––––––
Accuracy class* ––––––––––––––––––– k ––––––––––––––––––
Maximum capacity Max* g 5.1 2.1
Actual scale interval d* µg 1 0.1
Verification scale interval e* mg ––––––––––––––––––––1––––––––––––––––––––
Minimum capacity Min* mg 0.1 0.01
Tare range (subtractive) g –––––––––––––– < 105% of Max––––––––––––––
Response time (average) s –––––––––––––––––– 10 ––––––––––––––––––
Adaptation to ambient conditions
and application requirements –––By selection of 1 of 4 optimized filter levels –––
Display update (depends on the
filter level selected) s 0.1– 0.4 0.2–0.4
Allowable operating temperature range
– when the fully autom. self-calibr. funct.
“isoCAL”is on °C –––––––––––––––– 0 ... +40 ––––––––––––––––
– when the fully autom. self-calibr. funct.
“isoCAL”is off °C ––––––––––––––– +15 ... +25 –––––––––––––––
Pan size mm Ø 30 Ø 20
Height of draft shield mm –––––––––––––––––––122–––––––––––––––––––
Net weight, approx.
– Weighing cell kg –––––––––––––––––––3.3–––––––––––––––––––
– Evaluation system kg ––––––––––––––––––– 3.1–––––––––––––––––––
Power requirements V~ TNG6 portable power supply,
230 or 115,+15%...–20%
Frequency Hz –––––––––––––––––5060 –––––––––––––––––
Power consumption, incl. power supply,
approx. VA ––––––––18.7: maximum; 11: average ––––––––
Selectable weight units ––––––––––– Milligrams and grams –––––––––––
Selectable application programs Mass unit conversion; “EUREKA” air buoyancy
correction; differential weighing and
backweighing; diameter determination; tare
memory; net total; weighing in percent; over/under
checkweighing (classification and filling); counting
Automatic zero tracking device,
can be turned off by menu code –––––––––––––––– Standard ––––––––––––––––
Built-in interface RS-232 C-S/V24-V28, RS-423/V 10; 7-bit; parity:
even, odd; transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
* = EC Directive No. 90/384/EEC for non-automatic weighing instruments applicable in the
EC/EU Member States
5–5