Operating instructions

Weighing in the IQ-Mode*
(Load-Dependent Readability)
In the IQ-mode, weighing is done with a load-
dependent readability of, e.g., 0.01% (for different
settings, see part 2, “Balance Operating Menu”)
throughout the entire weighing range of the balance.
Oftentimes, a display accuracy of 10 mg
is sufficient for a load of approx. 110 g. In this case,
it makes sense to select weighing range R 1 with
an accuracy of 0.01% by pressing the w key (29).
While you are filling up to a target weight, it is
certainly easier to work with a target of 110.20 g than
with an absolutely accurate readout of 110.19885 g.
* = not with verified models; other settings in
the balance operating menu are necessary
on model MC 5
1–28