Operating instructions

Calibration Test
The calibration test is carried out with internal weights
as follows:
– MC 210 S and MC 210 P (-0CE)
with approx. 170 g
– MC 410 S (-0CE) with approx. 300 g
– MC 5 (-0CE) with approx. 5 g
– SC 2 (-0CE) with approx. 2 g
Press the tare key (26) for at least 2 seconds until
“C.I. and “CAL” are displayed (next to the f key).
Select the calibration test by pressing the F key (22)
twice until “C.t. is displayed.
Unload the balance and tare (close the draft shield).
C.t. stands for “calibration test.
When the balance displayes a zero readout, press
the f key (21). The built-in calibration weights are
now internally applied by servomotor. Afterwards, the
deviation of the momentary weight readout from the
target weight (displayed in grams only) is indicated.
If external interference affects the calibration test
procedure, you may obtain a brief display of the error
message “Err 02. In this case, tare and then press
the f key again when a zero readout is displayed.
The balance should be adjusted (calibrated) if the devi-
ation in the weight measurement no longer meets the
accuracy requirements of your particular application.
f key: The balance is automatically calibrated
(see also page 1–31)
or
F key: Quits the calibration test
An acoustic signal indicates the end of the sensitivity test.
Important Note:
For the setting “CAL calibration test using the f key”
see part 2, “Balance Operating Menu.
1–34