Operating instructions

Calibration Functions on Verified Balances Approved
for Use as Legal Measuring Instruments
Select the appropriate menu code to define the access status for each
of the following calibration functions, which are activated by holding down the
t key for a few seconds:
– External calibration C.E.
– Internal calibration C.I.
– Calibration C.t.
– Internal linearization L.I.
– Air density determination A.d. (see part 3, “Application Programs”)
However, if the menu access switch is unlocked (“accessible” status indicated
by
-C-
after you have accessed the balance operating menu), the
“external calibration” function will be accessible even though you have set the
menu code 1 9 2 for “access denied.
External calibration Code**
Accessible 1 9 1
Access denied 1 9 2*
Internal calibration Code
Accessible 1 10 1*
Access denied 1 10 2
Multiple Calibration Mode
The calibration value is calculated from the average of the individual calibration
procedures. You can use the “multiple calibration mode” for both internal
and external calibration. The number of calibration procedures is indicated in the
3rd place of the application display field (e.g., “C.I.3”).
Important Note:
If “Err 04” is briefly displayed, the calibration values deviate too much
from one another. This means the calibration value measured is not stored,
and the calibration procedure will be repeated.
Multiple calibration mode Code
Off 1 14 1*
On 1 14 2
** = factory setting
** = setting does not apply to verified balances approved for use as legal
measuring instruments
2–16