Operating instructions

The printout of the data record can have the following lines:
---------------------
MC1 - Sartorius : Balance family and manufacturer
Model MC5 : Balance model
S/N 030800046 : Serial no.of the balance
Id 4-32-1 : ID no.(e.g., for identifying
a workstation/operator)
---------------------
Date : 30-Jul-95 : Current date
Start: 10:05:30 : Time at which the application started
Ser. : : Field for entering the project number
Lim + 1.0 % : Reference data (in this case, “tolerance
and target weight”see also part 3,
Setp + 42.903 mg : “Application Programs”)
pRef + 100 %
Wxx% + 42.903 mg
Prc + 100.6 % : Value measured (in this case,
“calculated percentage”)
End : 10:05:45 : End of application
Name : : Field for signature of the operator responsible
---------------------