Operating instructions

Information and Printouts or Data Transfer
The info key i has two functions:
– one brief press: displays information and prints/transfers stored data;
– one long press: selects the reference percentage or reference sample
quantity for weighing in percent, diameter determination
or counting; clears the memory in the backweighing
mode (for more information, see the descriptions
of these programs)
Briefly press the i key to select the information
function: an “i” will appear on the display.
If you press f or F now, the data stored
in the memory will be retrieved and displayed. What
is displayed depends on the program selected.
Example: counting
Readout: reference sample quantity (pcs)
These data will also be output via the
nRef + xxxxxxxx pcs interface (printed or transferred) if you
wRef + xxxxxxxx g press the print key after the info key.
In this case, the print symbol P will be displayed
Data output only for next to the “i” until you press
code 7 2 2 a function key.
Data ID Code K* or NUM
Just by pressing a key on the balance, you can have numeric entries with ID codes
transmitted to an on-line device via the interface.
Function: Keys:
Output data ID code “K*” numeric entry + i
Output data ID code “NUM” numeric entry + p
Info + print + function key displays and prints data stored
Info + function key displays data stored in the memory
3–4