Operating instructions

Activating the Air Density Determination Function
Hold down the tare key for more than 2 seconds until
C.I.” and “CAL” are displayed next to the f key.
To activate the air density determination function,
press F, identified by “v,” several times until “A.d.
is displayed.
A.d.” stands for “air density.
Unload the weighing pan and tare the balance.
Starting Air Density Determination
To start air density determination, press the f key
when the display shows a zero readout. Afterwards,
a weight for steel “St” is displayed.
”Err 02”: The effects of an external disturbance
can cause the error code “Err 02” to
be displayed briefly. If this occurs,
tare the balance and then press the f
key again when the display shows
a zero readout.
Interrupt
procedure: You can stop an air density
determination procedure
at any time:
– By pressing the c key so that
a previously stored air density
is not changed or
– By pressing F identified in the
display by “RES” to set the standard
air density 1.2 kg/m
3
3–11