Operating instructions

– Aluminium Weight –
Use the numeric keys to enter the exact mass value
specified for the certified aluminum weight
(e.g., 5.018072), and store this value by
pressing the f key identified in the display by “STO.
If you press f, identified by “STO,” without
making a numeric entry, the display
changes to the density readout. Now you
can change it for a special aluminum weight.
Enter the mass value and press f to define
this density.
Center the aluminum weight on the weighing pan and
close the draft shield.
After the aluminum weight has been confirmed,
the calculated air density between 1.1 kg/m
3
and
1.3 kg/m
3
will be displayed.
If the air density determination result is
a value which is not between 1.0 kg/m
3
and
1.4 kg/m
3
, the determination procedure
is interrupted with two beeps and the
previous air density remains stored. If this
occurs, restart the air density determination
procedure after the two beeps.
Storing the Air Density and Ending the Procedure
– Store the air density displayed by pressing the
F key identified by “STO” or
– Reset the display to the standard value 1.2 kg/m
3
with the c key and store this value by pressing
the F key identified by “STO”
An acoustic signal indicates the end of the air density
determination procedure.
3–13