Figure 5-16. Dionex ICS-1100 Column Heater Cover Removed
d. Remove the columns from the column heater.
e. Pull the column heater straight toward you to remove it from the
component panel.
4. Before installing the new column heater, write down the two calibration
values (offset and slope) recorded on the back of the heater.
5. Align the connector on the back of the new column heater (P/N 069564) with
the connector on the component panel (see Figure 2-4
) and push the column
heater onto the component panel.
6. Remove the heater cover and the metal plate on the new heater.
7. Install the columns in the new heater and replace the top metal plate and
heater cover.
8. Turn on the Dionex ICS-1100 power.
9. Enter the column heater calibration values in Chromeleon as follows.