Operator`s manual

Table Of Contents
A Specifications
Doc. 065289-03 10/12 123
A.8 Conductivity Cell with Heat Exchanger
A.9 Injection Valve
Full-Scale
Output Ranges
Digital signal range: 0 to 15,000 S
Analog signal range: 0 to 15,000 S
Electronic Noise ±0.1 nS/cm when background conductivity is 0 to 150 S/cm
±2 nS/cm when background conductivity is 151 to 3200 S/cm
Filter Rise times from 0 to 10 s; user-selectable
Temperature
Compensation
Fixed at 1.7% per 1 °C at cell temperature
Temperature
Range
Ambient +7 °C; 30 to 55 °C
Temperature
Stability
<0.01 °C
Cell Body Chemically inert polymeric material
Cell Electrodes Passivated 316 stainless steel; compatible with MSA
Cell Volume <1 L
Maximum Cell
Operating
Pressure
10 MPa (1500 psi)
Heat Exchanger Inert, tortuous path for low axial dispersion
Operating
Temperature
30 to 55 ºC (86 to 130 ºF); must be at least 7 ºC above the ambient
temperature
Injection Valve 6-port, 2-position Rheodyne valve with PEEK wetted components;
electrically-activated