Doc. 065289-03 10/12 17
2.1.4 Right-Side Panel
Figure 2-5 illustrates the RFIC-ER compartment on the right side of the
Dionex ICS-1100.
The compartment houses the ER Controller (ERC 10) and various
columns required for operation in the RFIC-ER mode. The ERC 10
electronics monitor the volume of electrolysis gases (such as hydrogen
and oxygen) in the eluent stream. If the predefined safety limit is
exceeded, the ERC 10 shuts down the pump to prevent the buildup of
gases in the eluent bottle.
If you plan to operate in the RFIC-ER mode, you must order one of these
kits:
RFIC-ER Anion Startup Kit, ICS-1100/ICS-1600 (P/N 069570)
RFIC-ER Cation Startup Kit, ICS-1100/ICS-1600 (P/N 069569)
Each Startup Kit includes one Installation Kit and one Consumables Kit.
Together, these kits contain additional items (purification columns, a
Figure 2-5. RFIC-ER Compartment (Cover Removed)