Flow Description for RFIC-ER Mode
Eluent from the eluent reservoir flows first through the pump degas
assembly (if it is installed) and then through the eluent valve
to the pump.
The eluent is then pushed through the pressure transducer
, which measures
the system pressure.
From there, the eluent is pushed through a pulse damper, which smooths
minor pressure variations from the pump to minimize baseline noise. The
eluent then flows through the ER1 column
and into the injection valve .
After sample is loaded into the sample loop and the injection valve is
toggled to the Inject position, eluent passes through the sample loop.
If the optional column heater is installed, the eluent/sample mixture is
pumped through the heat exchanger
, which heats the mixture to the column
heater temperature.
The mixture then goes to the guard and separator columns and through the
suppressor .
From the suppressor, the mixture flows through the cell , where the analytes
are detected. A digital signal is sent to Chromeleon software. Analog output
can be collected simultaneously.
Finally, the mixture flows out of the cell, through the ER2 column , and is
then recycled back into the suppressor.
From the suppressor, flow moves to the ER3a and ER3b columns, which
remove contaminants from the eluent.
The ERC 10 monitors the eluent for gas and shuts down the pump if the gas
exceeds a pre-defined limit.
Finally, the output from the ERC 10 is recycled back to the eluent bottle.