26 Doc. 065289-03 10/12
2.4 System Component Details
This section provides details about Dionex ICS-1100 system components,
including the vacuum degas assembly (optional), RFIC-ER components, eluent
valve, pump, injection valve, column heater (optional), suppressor, and
conductivity cell.
2.4.1 Vacuum Degas Assembly (Optional)
The vacuum degas assembly provides online eluent degassing at the time
and duration specified by the user. The assembly, which must be installed
in the Dionex ICS-1100 at the factory, consists of:
A single-channel degas chamber (with degas membranes) with
internal capacity of 17 mL
A dual-stage diaphragm vacuum pump
A solenoid valve
An on-board vacuum sensor
The electronics required to operate the vacuum pump
Tubing, fittings, and other accessories
By default, the Dionex ICS-1100 monitors the degas pressure reading and
turns the degas pump on and off as needed. Different degas operating
modes can be selected from Chromeleon.
To select the degas operating options, open the Chromeleon Server
Configuration program, right-click the Dionex ICS-1100 device in the
timebase, and select Properties.