Operator`s manual

Table Of Contents
3 Operation and Maintenance
Doc. 065289-03 10/12 49
3.10 Loading and Injecting Samples
NOTE Samples can be injected using either the standard
injection valve or the optional auxiliary valve, if
installed. For more information, see the Chromeleon
Help or user’s manual.
The two techniques for loading samples into the sample loop are: (1) with an
autosampler or (2) with a syringe or vacuum syringe through the injection port on
the Dionex ICS-1100 front door.
For autosampler injections, the injection port tubing is disconnected from the
Dionex ICS-1100 injection valve and replaced by the autosampler outlet tubing.
Other setup requirements depend upon the model of autosampler.
Setup for a Dionex AS Autosampler
To use a Dionex AS Autosampler with the Dionex ICS-1100, the following
requirements must be met:
The Dionex ICS-1100 injection valve (or the auxiliary valve, if installed) must
be specified as the injection valve and must be linked to the Dionex AS in the
Chromeleon timebase.
An injection valve should not be installed in the AS.
See the Dionex ICS-1100 installation instructions for Dionex AS connection
instructions.
Setup for a Dionex AS-DV Autosampler
To use a Dionex AS-DV Autosampler with the Dionex ICS-1100, the following
requirements must be met:
The Dionex ICS-1100 injection valve must be linked to the Dionex AS-DV in
the Chromeleon timebase.
Setup for a Dionex AS-HV Autosampler
For setup information for the AS-HV Autosampler, see the AS-HV Autosampler
Operators Manual (Document No. 065125). The manual is provided on the
Thermo Scientific Reference Library DVD (P/N 053891).