Dionex ICS-1100 Ion Chromatography System
50 Doc. 065289-03 10/12
3.10.1 Loading Samples with a Syringe
1. Make sure the injection port on the Dionex ICS-1100 front door (see
Figure 2-1
) is connected to sample port S (5) on the injection valve
(see Figure 3-7
).
2. Fill the 1 cc syringe (P/N 016388) provided in the Dionex ICS-1100
Ship Kit (P/N 057905) with a calibration standard or sample.
3. Insert the syringe into the injection port on the Dionex ICS-1100 front
door (see Figure 2-1
).
4. Verify that the injection valve is in the Load position.
5. Overfill the sample loop with several sample loop volumes. Excess
sample will exit through the injection valve waste line.
6. Leave the syringe in the port.
7. Switch the injection valve to the Inject position (see Section 3.10.4
).
Figure 3-7. Injection Valve Connections