Operator`s manual

Table Of Contents
4 Troubleshooting
Doc. 065289-03 10/12 61
4.2 Troubleshooting Error Messages
This error occurs when the column heater temperature is higher than the
maximum allowed. This error may occur if the Dionex ICS-1100 is operating
in an extreme environment (greater than 40
°C (104 °F)).
To troubleshoot:
Refer to Section A.3
for environmental specifications.
This error occurs when the column heater is unplugged from the component
panel.
To troubleshoot:
1. Check that the column heater is plugged into the component panel.
2. If the error persists, the column heater may be faulty. Replace the column
heater (see Section 5.12
).
This error occurs when there is a short circuit on the thermistor input used to
measure the column heater temperature. This error indicates a faulty column
heater.
To troubleshoot:
Replace the column heater (see Section 5.12
).
This error occurs when the degas vacuum does not reach the expected level
during calibration.
COLUMN HEATER EXCEEDS SAFE TEMPERATURE
COLUMN HEATER OPEN CIRCUIT
COLUMN HEATER SHORT CIRCUIT
DEGAS CALIBRATION FAILED