This error indicates a problem in the pump controller electronics.
To troubleshoot:
Contact Thermo Fisher Scientific for assistance. The Dionex ICS-1100
electronics components cannot be serviced by the user.
If the system pressure exceeds the set limit for at least 0.5 second, this error
message appears and the pump stops. Pressure limits can be set in the
Chromeleon Server Configuration or in the control program.
To troubleshoot:
1. Check for blockages in the liquid lines by working your way backward
from the cell to the pump. See Section 2.2
for the system flow schematics.
2. Check that the flow rate is set to the correct value.
3. Check that the high pressure limit is set to the correct value.
4. Restart the pump.
This error occurs if the Dionex ICS-1100 loses communication with
Chromeleon.
To troubleshoot:
Verify that the USB cable is connected correctly from the Dionex ICS-1100
rear panel to the Chromeleon PC.
PUMP PRESSURE HARDWARE ERROR
PUMP OVER PRESSURE
PUMP STOPPED DUE TO LOST USB COMMUNICATION
ERROR