4 Troubleshooting
Doc. 065289-03 10/12 65
If the system pressure falls below the low pressure limit, the pump stops and
this error message appears. Pressure limits can be set in the Chromeleon
Server Configuration or in the control program.
To troubleshoot:
1. Make sure the eluent reservoir is full.
2. Check for liquid leaks (see Section 4.3
).
3. Check that the waste valve is closed.
4. Prime the pump (see Section 5.16
).
5. Restart the pump.
If the auxiliary valve fails to switch position within 1 second of being toggled,
the Dionex ICS-1100 Moduleware reports an error to Chromeleon and this
error message appears.
To troubleshoot:
1. If a sequence is being executed, terminate the sequence by selecting
Batch > Stop in Chromeleon.
2. Try to toggle the valve from position A to position B by clicking the
corresponding buttons on the Dionex ICS-1100 Control panel in
Chromeleon (or by clicking the Valve_2 commands under Pump_ECD in
the Commands dialog box).
3. Turn off the Dionex ICS-1100 power briefly and then restart.
4. If the problem persists, repeat Step 2
.
5. If the problem persists, contact Thermo Fisher Scientific for assistance.
PUMP UNDER PRESSURE
SECOND VALVE ERROR