This error occurs if you turn on the suppressor and the Dionex ICS-1100
cannot establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see Section 5.11
).
2. If the error persists, contact Thermo Fisher Scientific for assistance. There
may be a problem in the suppressor controller electronics. The Dionex
ICS-1100 electronics components cannot be serviced by the user.
This error may be caused by an expended or dirty suppressor or by a
malfunction in the suppressor controller electronics.
To troubleshoot:
1. Follow the instructions in the suppressor manual to regenerate the
suppressor. Suppressor manuals are included on the Thermo Scientific
Reference Library DVD (P/N 053891).
2. Follow the instructions in the suppressor manual to clean the suppressor.
3. If a malfunction in the suppressor controller is suspected, contact Thermo
Fisher Scientific for assistance. The Dionex ICS-1100 electronics
components cannot be serviced by the user.
SUPPRESSOR NOT CONNECTED
SUPPRESSOR OVER CURRENT