Operator`s manual

Table Of Contents
4 Troubleshooting
Doc. 065289-03 10/12 69
Leaking pressure transducer
1. Make sure the liquid line connections into the pressure transducer are
tight. Refer to Installation of Dionex Liquid Line Fittings (Document No.
031432) for tightening requirements. The manual is included on the
Thermo Scientific Reference Library DVD (P/N 053891). Replace any
damaged fittings.
2. If the pressure transducer continues to leak, contact Thermo Fisher
Scientific for assistance. The pressure transducer cannot be serviced by
the user.
Leaking pump head waste valve
Make sure the waste valve is closed. To close the valve, turn the knob
clockwise, and tighten fingertight. DO NOT OVERTIGHTEN!
Overtightening may damage the valve and the pump head.
If the leak persists, replace the waste valve O-ring (see Section 5.9
).
Leaking suppressor
Refer to the suppressor manual for troubleshooting procedures. Suppressor
manuals are included on the Thermo Scientific Reference Library DVD
(P/N 053891).
Leaking injection valve or auxiliary valve
1. Make sure the liquid line connections to the transducer are tight. Refer to
Installation of Dionex Liquid Line Fittings (Document No. 031432) for
tightening requirements. The manual is included on the Thermo Scientific
Reference Library DVD (P/N 053891). Replace any damaged fittings.
2. Liquid leaks from behind the valve stator may indicate a scratched rotor
seal. Rebuild the valve (see Section 5.4
).
Leaking detector cell
1. Check the waste lines for blockage; trapped particles can plug the lines
and cause a restriction and/or leak. If necessary, clear the waste lines by
reversing the direction of flow.
2. Make sure the plumbing downstream from the cell is clear; a blockage
may overpressurize the cell, causing it to leak. If the problem continues,
contact Thermo Fisher Scientific for assistance.